HYSTERESIS
1) SINGLE CELL
SCALES:
a. take
a tare - reading to tenths of a mv.
b. put cell under tension -- remove load, note tare in mv.
c. put cell in compression -- remove load, note tare in mv.
d. Tare readings must return to original within 0.1 mv.
If not, cell is damaged, or the scale bearings are bad.
2)
MULTIPLE CELL SCALES:
ISOLATE signal from cells, then follow as in single cell systems except take
readings for each cell.
To isolate signals from paralleled cells:
A. Determine the return signal wire colors.
Colors for
loadcells: REVERE/H.B.M. Most Others
+ EXCITATION to
cell Green Red
- EXCITATION to
cell Black
Black
+ SIGNAL from cell in
tension Red White
- SIGNAL from
cell in tension
White Green
B. Separate the signal wires at the junction
box.
C. Verify that the excitation is still
applied to each cell.
D. Apply excitation (system on), verify 8-15 volts d.c.
(typical)
E. Write down empty scale tare milli-volt
readings from each cell.
F. Add weight to scale progressively, write down readings for
each cell.
Where possible,
apply stresses at cell under test.
LINEARITY
These examples assume 30mv/volt loadcells with 10 volt
excitation.
1)
SINGLE CELL SCALES
should progress evenly in
signal value returned for load applied
i.e. this example shows readings
from a single cell scale with a 1500#
loadcell:
LOAD SIGNAL COMPUTED WEIGHT ON CELL
Tare Only: 12
mv. 12/30mv. x 1500# = 600#
tare
300#
18mv. 18/30mv. x 1500# = 900#
600#
24mv. 24/30mv. x 1500# = 1200#
900#
30mv. 30/30mv. x 1500# = 1500#
2)
MULTIPLE CELL SCALES,
example of a 3-cell scale
using 3 x 500# cells and a scale with a 600#
tare.
LOAD CELL-1 / weight CELL-2 / weight CELL-3
/ weight
Tare: 12 mv.
200# 11.4 mv. 190# 12.6 mv. 210#
300# 18 mv.
300# 17.1 mv. 285# 18.9 mv. 315#
600# 24
mv. 400# 22.8
mv. 380# 25.2 mv. 420#
900# 30
mv. 500# 28.5 mv. 475# 31.5 mv.
525#
NOTE: SIGNALS were
measured, WEIGHTS were computed and should total up to actual, i.e. Tare =
200 + 190 + 210 = 600#.
EVALUATION OF CELL DATA FOR SCALE ACCURACY
This can be
best summarized by documenting the working range of the cell for its
LOADING.
1) SINGLE
LOADCELL Scales:
Used (for
example) with 10 volts excitation and a 3 mv. /volt response (0mv. no-load,
30mv. full load), should be tested for tare and full scale signal readings.
- If the full-scale minus tare value is less than 20% (6mv.), gain and
accuracy will be impaired.
- A good
solution is to mechanically balance out the tare in the scale so that a
smaller cell can be used.
2) MULTIPLE
LOADCELL Scales:
The cells
should be isolated from one another in the return signals and be tested as
in the single cell applications for tare and loaded readings.
- The cells should read within 5 mv. of each other
or the cell placement should be examined for even load distribution.